공고마당 로고공고마당
공고원문다운로드

COMMODITY DESCRIPTION

H.S. NOITEM NO.DESCRIPTIONQUANTITYUNIT
Electrocardiograph, analysis
software
1system

A . FEATURES :

1. The software that analyzes electrocardiogram images or waveform data using deep learning algorithms

to assist in the diagnosis of arrhythmias and provides numerical values ​​for the risk of emergency

situations and cardiac dysfunction, thereby assisting doctors in making diagnostic decisions.

2. 11 types of heart rhythms can be assessed (Sinus Rhythm, Atrial Fibrillation, Atrial Flutter, Paroxysmal

Supraventricular Tachycardia (PSVT), Atrial Rhythm, Frequent Premature Beats, Pacemaker Rhythm,

Pre-Excitation, Ventricular Tachycardia, Second-degree AV Block, Third-degree AV Block).

3. 10 digital biomarker disease risk assessments are available (Critical Condition (risk of shock, cardiac

arrest, and respiratory failure within 24 hours), Acute Coronary Syndrome (ACS), Acute Myocardial

Infarction (STEMI), Myocardial Injury, and Left Ventricular Dysfunction (LVD). Dysfunction), right

ventricular dysfunction (RV Dysfunction), pulmonary edema, pulmonary hypertension, large pericardial

effusion (cardiac tamponade), hyperkalemia

B. CONSIST OF :

1. Electrocardiograph, analysis software 1 set

- Gateway Server

2. Warranty (same as the subcription service contract)

3. Operating manual 1 copy

C. SPECIFICATIONS :

1. Electrocardiograph, analysis software

1) Data input : 12 lead ECG (image or waveform data)

: Information of user input could be verified

2) Presentation of types and probability of arrhythmia

- Sinus rhythm (tachycardia, bradycardia, arrhythmia, sinoartrial block)

- Atrial fibrillation (slow vetricular response, regularized fib., preexcitation)

- Atrial flutter

- PSVT

- Atrial rhythm (focal atrial tachycardia, multifocal atrial tachycardia, wandering atrial pacemaker,

atrail bradycardi)

- Frequent premature beats (supraventricular bigeminy/trigeminy, frequent premature ventricular

contractions)

- Electronic pacemaker

- Pre-excitation (atrial fibrillation with pre-excitation)

- Ventricular tarchycardia (non-sustained ventricular tarchycardia)

- Second-degree block (atrioventricula block-Wenckebach/Mobitz type 2,

atrioventricula block(2:1)/high grade)

- Thrid-degree bolck (complete heart block)

3) Presentation of scores and confidence intervals for digital biomarkers (risk of emergency

Page 1 of 3 신청부서: 의공기술검토자: 성 기 웅

situations and cardiac dysfunction)

- Ciritical condition : Risk of shock/respiratory failure/cardiac arrest within 24 hours

- ACS, STEMI, Pimlonary edema, left/right ventricular dysfunction, pulmonary hypertension

- Myocardial injury, Large pericardial effusion (or tamponade)

- Severe hyperkalemia ( K 6.0 mEq or more)

4) AUC (minimum)

I. Arrhythmia

- Sinus rhythm: 0.899

- Atrial fibrillation: 0.912

- Atrial flutter : 0.882

- PSVT : 0.886

- Frequent prematur beat : 0.827

- Atrial rhythm : 0.815

- Electronic pacemaker : 0.867

- Pre-excitation : 0.746

- Ventricular tarchycardia : 0.794

- 2nd degree block : 0.833

- 3rd degree block : 0.860

II. Digital biomarker

- Ciritical condition : 0.760

- ACS : 0.755

- STEMI : 0.917

- Myocardial injury : 0.786

- Pulmonary edema : 0.827

- Large pericardial effusion : 0.810

- Left ventricular dysfunction : 0.761

- Right ventricular dysfunction : 0.802

- Severe hyperkalemia : 0.850

5) Gateway server

Limited to equipment specifications that have passed the security certification by the KNUH

-

Information Security Team

- Type : RACK

- Name : HPE DL20 Gen11 E-2434 3.4GHz 4C 1P 16GB-U 2LFF 290W PS Svr

(E-2434 4core 3.4GHz 1P, 16GB x1, VROC, 290W)

- Processor : Intel Xeon E-2434 3.4GHz 4-core 55W Processor

- Memory : HPE 16GB (1x16GB) Single Rank x8 DDR5-4800 CAS-40-39-39 Unbuffered

Standard Memory

- Network Controller : HPE embedded 1GB 4-port BCM5719 Network Adapter

- Storage Controller : embedded Intel VROC SATA software RAID

- Internal Stroage : 2 LFF HDD Bays (Hot Plug)

- Optical Drive Bay : 1; (Optional: DVD-ROM, DVD-RW)

- Management : iLO Management (standard), Intelligent Provisioning (standard)

- Power Supply : HPE DL20 Gen10 290W Platinum Power Supply Kit (2 LFF Only)

D. REMARKS :

1. Installation

Page 2 of 3 신청부서: 의공기술검토자: 성 기 웅

- It shall be provided free of charge during the contract period, and installation and operation must be

supported at a location designated by KNUH.

- The vendor is solely responsible for all tasks and costs related to the installation, initial setup, and

normal operation of the server.

2. Operation and Maintenance Requirements

- Provide PACS integration and technical support free of charge.

- Provide updates and security patches for performance improvement during the contract period.

- Maintain a 24-hour operational failure response system and provide remote or on-site technical support

in the event of a failure.

Page 3 of 3 신청부서: 의공기술검토자: 성 기 웅

AA