Automatic Genetic Analyzer

(Human Identification)

set

1

A. 용도(End-user’s Use)

1. The system offers continuous, unattended operation with automated polymer loading with automated polymer delivery system, sample injection, separation and detection, and data analysis.

2. With 24 capillaries, hundreds of sequencing and fragment analysis samples can be analyzed in a day.

3. Advanced thermal system design improves temperature control for demanding DNA fragment analysis application.

4. The standardization function of the data result signal can correct the deviation of the result caused by the sample and the test condition and environment difference.

5. The system offers key consumables data and record administrative information with Radio Frequency Identification (RFID) technology such as remaining volume, available run count, and expiration date etc.

6. Advanced multiplexing capabilities for DNA fragment analysis with six or more unique dyes.

B. 장비의 구성(Configurations of Good’s)

I) Mainbody

1. Automatic Genetic Analyzer ļ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 system

2. Data Control System ු Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

3. Sequencing Analysis Software സ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

4. GeneMapper ID-X Software Ɣ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

II) Sub unit

1. Accessories 8 Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

C. 성능 및 규격(Performance and Specification) (Configurations below are based on 1 set)

I) Mainbody

1. Automatic Genetic Analyzer ļ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 system

1) Capillary Electrophoresis

(1) With more efficient heat dissipation than slab gels, capillaries permit higher run voltages, resulting in faster run times.

(2) Capillary electrophoresis enables the use of electrokinetic injection to load samples into the capillaries.

(3) Samples are simultaneously injected into the parallel 24 capillary array in less than 30 seconds.

(4) The most dramatic advantage of capillary electrophoresis is the elimination on manual operations, resulting in increased run-to-run consistency and reliability.

(5) Electrophoresis Voltage : Up to 20 kV.

2) Detection System

(1) The emitted fluorescent light is collected, separated by wavelength and focused onto a charged-coupled device (CCD) or equivalent.

(2) The fluorescent light is collected and dispersed across the CCD or equivalent, the data is transferred to the instrument computer and then undergoing chemometric algorithmic processing, the data must be capable of being transformed into up to six or more electropherograms.

3) Laser : Long-life, single-line 505 nm, solid-state laser excitation source.

4) Oven Temperature : Active temperature control from 18℃ to 70℃.

5) System Performance and Throughput

(1) Rapid Sequencing Run Module

① Performance : Median bases collected in 90% of samples ≥600

② Polymer type : Performance Optimized Polymer 7

③ Capillary : Ld = 50cm

④ Average run time : ≤40 minutes

⑤ Average throughput/day : 840 samples, 35 runs, 420,000 bases/day

(2) Fast Sequencing Run Module

① Performance : Median bases collected in 90% of samples ≥750

② Polymer type : Performance Optimized Polymer 7

③ Capillary : Ld = 50cm

④ Average run time : ≤65 minutes

⑤ Average throughput/day : 504 samples, 21 runs, 352,800 bases/day

(3) Standard Sequencing Run Module

① Performance : Median bases collected in 90% of samples ≥1,000

② Polymer type : Performance Optimized Polymer 7

③ Capillary : Ld = 50cm

④ Average run time : ≤125 minutes

⑤ Average throughput/day : 240 samples, 11 runs, 204,000 bases/day

(4) ShotRead Sequencing Run Module

① Performance : Median bases collected in 90% of samples ≥325

② Polymer type : Performance Optimized Polymer 7

③ Capillary : Ld = 50cm

④ Average run time : ≤30 minutes

⑤ Average throughput/day : 1,104 samples, 48 runs, 331,200 bases/day

(5) Fragment Analysis Run Module

① Resolution range in ≥90% of samples : ≤40 to ≥520

② Largest fragment collected in ≥90% of samples : ≥600

③ Sizing precision of 100% of alleles in ≥90% of samples : 50-400 bp - <0.15, 401-600 bp - <0.3

④ Multirun sizing of 100% of alleles in ≥90% of samples : 50-400 bp - <1 bp, 401-600 bp - <2bp

⑤ Polymer : Performance Optimized Polymer 7

⑥ Capillary : Ld = 50cm

⑦ Average run time : 40 minutes

⑧ Ǵ Ā Average throughput/day : ≥840 samples

(6) Long fragment Analysis Run Module

① Ǵ Ā Resolution range in ≥90% of samples : ≤40 to ≥700

② Ǵ Ā Largest fragment collected in ≥90% of samples : ≥1,200

③ Ǵ Ā Sizing precision of 100% of alleles in ≥90% of samples :

50-400 bp - <0.15, 401-600 bp - <0.3, 601-1,200 bp - <0.45

④ Ǵ Ā Multirun sizing of 100% of alleles in ≥90% of samples :

50-400 bp - <1 bp, 401-600 bp - <2bp, 601-1,200 bp - <3bp

⑤ Ǵ Ā Polymer : Performance Optimized Polymer 7

⑥ Ǵ Ā Capillary : Ld = 50cm

⑦ Ǵ Ā Average run time : 125 minutes

⑧ Ǵ Ā Average throughput/day : ≥264 samples

(7) SNaPshot Analysis Run Module

① Ǵ Ā Resolution range in ≥90% of samples : ≤40 to ≥120

② Ǵ Ā Largest fragment collected in ≥90% of samples : ≥120

③ Ǵ Ā Sizing precision of 100% of alleles in ≥90% of samples : 50-400 bp - <0.5

④ Ǵ Ā Multirun sizing of 100% of alleles in ≥90% of samples : 50-400 bp - <1 bp

⑤ Ǵ Ā Polymer : Performance Optimized Polymer 7

⑥ Ǵ Ā Capillary : Ld = 50cm

⑦ Ǵ Ā Average run time : 30 minutes

⑧ Ǵ Ā Average throughput/day : ≥1,104 samples

6) System Consumables

(1) Capillary Arrays

① The internally uncoated capillaries are supplied in assemblies of 24 capillaries per array, with a built-in frame for easy installation.

② Ǵ Ā Arrays are avaiable in 36cm and 50cm capillary lengths to support multiple applications and run methodologies.

③ Ǵ Ā The 50cm array has been optimized for many applications for use with POP-7 polymer and specific run modules.

④ Ǵ Ā Capillary i.d. : 50µm.

⑤ Capillary lifetime : last for more than 160 runs

(2) Performance Optimized Polymers

① The polymer must be provided in sealed pouchs to prevent contamination and be available in 3 pouch(POP-7, POP-6 and POP-4) configurations, enabling user to select the polymers type based on their experimental requirements.

② Ǵ Ā The polymers are offered to support the processing of 384 or 960 samples, depending on sample throughput requirements

③ Ǵ Ā The pouchs has adequate polymer to support the started number of samples or injections, plus additional volume for initial setup and installation operation.

④ Ǵ Ā Most applications may be run with POP-7 polymer, along with the 50 cm array.

(3) Buffers and Conditioning Reagent Consumables

① Ǵ Ā To minimize the risk of contamination during dilution, the product is supplied in a sealed format and is ready for immediate use upon opening, with no additional preparation required.

② Ǵ Ā The system must provide information related to consumable usage, including details such as expiration date.

(4) Radio Frequency Identification (RFID) Labeling

① The system incorporate RFID labels on the capillary arrays, polymer pouches, buffer containers, and conditioning pouches.

② Ǵ Ā The system provides information regarding consumable usage count, identification number, expiration date, and the number of samples available for analysis.

(5) Instrument Normalization Reagent

① The GeneScan 600 LIZ size standard v2 is specially formulated for dual-purpose use on the System.

② Ǵ Ā The reagent may be used as a size standard for DNA fragment sizing applications.

7) Option

(1) An optional upgrade provides security, audit, and eSig features to assist laboratory compliance with 21 CFR Part 11 requirements.

(2) 4 plates can be run at once.

2. Data Control System ಈ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

1) Maker supplied Dell Computer which is all tested and certified by manufacturer

2) Computer Specificaiton

(1) Hardware : Core i9-12900, 16 Cores/2.4GHz to 5.1GHz, 30MB

(2) Operating system : Windows 11 Enterprise, Media, 64-bit, English

(3) Installed RAM : 16 GB, Non-ECC, 2666MHz

(4) Hard Disk Storage : 2x512 GB M2 2230 PCIe NVME

(5) Monitor : LCD 27' Color display

3. Sequencing Analysis Software സ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

1) Data collection software

2) Communicates instruction to firmware on the instruments that controls

instrument function

3) Created and edit sample sheets and run files

4) Perform a run and view the resulting data

5) Sequence data analysis software

6) Analyzes the raw sequencing data collected by the data collection program and calls the base sequence

7) Perform a special tasks relating to gel files (Gel Doc)

8) Make and copy matrices (Data utility)

4. GeneMapper ID-X Software Ɣ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

1) Consolidates GeneScan/Genotyper software features into one convenient software package

2) Novel PQVs ensure allele-calling accuracy

3) Integrates data from all laboratory instruments for ease-of-use

4) Provides fast, simple, accurate data analysis to improve research methods

5) GeneMapper IDX software increased productivity six-fold compared to allele calling with GeneScan and Genotyper software.

6) The GeneMapper IDX software is a high-performance genotyping tool designed to eliminate the need for human intervention provide very high throughput and accuracy required by automated genotyping users.

7) A single mouse click process then carries out the entire analysis process.

8) Rather than returning a single quality value, GeneMapper software assigns a seriesof quality values that lets the user rapidly pinpoint the is for failed allelecalls.

9) The use of quality values substanitally minimizes the amount of time spent analyzing data.

10) 1 set is consist of Full version and Client version together.

11) GeneMapper IDX software is 99,99% accurate when comparing high quality calls with those made manually.

(1) Automated the entire process of separating, quantifying and sizing PCR and other DNA fragments using internal-lane standards

(2) Analyzes microsatellites fast and reliably

(3) Eliminates lane to lane or run to run variability and provides the most accurate DNA fragment sizing

(4) Flexible display of results : tabular form (elution time, fragment size, amount of fluorescence signals in peak, peak height, peak area, and the type of dye labeled)

(5) Printing and exporting data

II) Sub unit

1. Accessories 8 Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 set

1) Installation Consumables

(1) POP-4 (960) POLYMER 3500 SERIES ؘ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(2) CATHODE BUFFER CONTAINER 3500 SERIES Ѵ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(3) ANODE BUFFER CONTAINER 3500 SERIES ࣜ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(4) CONDITIONING REAGENT 3500 SERIES d Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 4 ea

(5) TF HI-DI FORMAMIDE 4 x 5mL BOTTLE ౸ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(6) FG GENESCAN-600 LIZ SIZE STD V2.0 ྀ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(7) FG,CAPILLARY ARRAY 24-CAP 36CM RUO ࣘ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(8) HID 6-DYE Installation Standard ത Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(9) DS-36 MATRIX STANDARD KIT (DYE SET J6) ׀ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

2) Data Backup System for ID(Identification) analysis ̨ Ā ྠ Ā ྠ Ā ྠ Ā ྠ Ā 1 ea

(1) Hardware : Core i9-12900, 16 Cores/2.4GHz to 5.1GHz, 30MB

(2) Operating system : Windows 11 Enterprise, Media, 64-bit, English

(3) Installed RAM : 16 GB, Non-ECC, 2666MHz

(4) Hard Disk Storage : 2x512 GB M2 2230 PCIe NVME

(5) Monitor : LCD 27' Color display

D. 비고(Remark)

1. The installation and operation should be carried out by bidder / supplier's responsibility at user's site. and If there are several installation site, installation should be carried out at each site

2. One year warranty service should be provided by supplies after the performance test.