공통 규격서 (SPECIFICATION)
H.S.No. Description Q'ty
3.0 Tesla MRI System 1 System
A. Features:
1. 3.0T actively-shielded magnet with high homogeneity.
2. Full range of whole body applications capability.
3. Fully channel independent RF receive system.
4. High resolution imaging with surface coils.
5. The dedicated multi channels coils for parallel imaging technique.
6. Reconstructor ensures fast, interactive and real-time images.
7. Advanced MR applications for Diffusion and Perfusion.
8. The powerful gradient system for the ultrafast imaging.
B. Configuration:
1. Magnet system 1 set
2. Gradient system 1 set
3. RF system 1 set
4. Patient support 1 set
5. Coil solutions 1 set
6. Host computer 1 set
7. Reconstruction system 1 set
8. Operator console 1 set
9. Standard software package 1 set
10. Advanced applications 1 set
11. MR Independent 3D Workstation System 1 set
12. Site Equipments 1 set
13. Accessories 1 set
14. Unique Items of each vendor : [Type A] or [Type B] 1 set
C. Specifications:
1. Magnet system 1 set
1) Field strength : 3.0 Tesla
2) Type: Compact super conducting magnet or equivalent
3) Bore diameter : 70cm or more
2. Gradient system 1 set
1) The highest performance of the gradient : XP or XT-Engine
2) Max. Gradient amplitude and Slew Rate (per axis) : 65mT/m & 220 T/m/s or 60mT/m & 200T/m/s
3) Max. b-value : 10,000 or more
4) Cooling Type : Direct liquid cooling or equivalent
5) Duty Cicle : 100% or equivalent
6) Max. FOV : 55cm or more
7) Min. FOV : 5mm or less
3. RF system 1 set
1) The highest performance of RF system : MultiTransmit RF or TimTX TrueShape RF
2) The number of independent RF amplifiers : 2 amps
3) Output power : 36kW or more
4) Amplitude resolution : 16bits or more
5) Number of system channel that can be used simultaneously
: Channel independent or 128 channels
6) Signal chain from coil to reconstructor : Fully digital or Direct RF
7) Signal chain from connector to magnet : Digital
8) Receive signal resolution : 32bits or more
4. Patient support 1 set
1) Max. weight capacity :250kg or more
2) Horizontal table speed : max. 200 mm/s or more
3) Table travel : 268cm or more
4) Patient headset
5) Physiological synchronization
6) Peripheral pulse unit
7) Respiration sensor unit
8) Connectors
5. Coil solutions 1 set
1) Ingenia T/R body coil or Quadrature RF Body Coil 1 ea
2) Head neck coil set : dS 20ele. HeadNeck coil or BioMatrix Head/Neck 20 coil 1 ea
3) Spine coil set : dS 44ele. TotalSpine coil or BioMatrix Spine 72 coil 1 ea
4) Body coil set : dS 32ele. Torso coil or Contour 24 coil 1 ea
5) Knee coil set : dS 16ele. Knee coil or TX/RX Knee 18 coil 1 ea
6) Foot Ankle coil set : dS 16ele. FootAnkle coil or Foot/Ankle 16 coil 1 ea
7) Shoulder coil set : dS 16ele. Shoulder coil or Shoulder Shape 16 coil 1 ea
8) Wrist coil set : ds 16ele. Hand/Wrist coil or Hand/Wrist 16 coil 1 ea
9) Wholebody coil : dS WholeBody coil (2nd dS Torso) or Contour XL coil 1 ea
10) Specialized coil set : dS 16ele. SmallExtremity coil or Breast 18 coil 1 ea
11) High performance head coil : dS 32ele. Head coil or BioMatrix Head/Neck 64 coil 1 ea
6. Host computer 1 set
1) Operating system
2) Processor : Ten Core Intel Processor or Intel Xeon (10Core)
3) Clock rate: 3.5GHz or more
4) Memory : 32 GB or more
5) Hard disk: 1000GB or more
6) Media drives : DVD or CD/DVD
7. Reconstruction system 1 set
1) Processor : 3.7 GHz Ten Core Intel processors or 2 x 16 core Intel Xeon gold 6226R
2) Memory: 96 GB or more
3) Reconstruction speed (256 x 256 FFT) : 142,222 recons/sec or more
8. Operator console 1 set
1) Display size : 27 inch x 2 or 24 inch x 2
2) Resolution : 1280 x 1024 pixels or more
9. Standard software package 1 set
1) Turbo Spin Echo
2) Single-shot TSE
3) TSE with 90 flip-back pulse
4) 3D TSE with variable flip angle
5) Number of echos in TSE
6) Gradient Echo
7) Spoiled gradient echo
8) Coherent gradient echo
9) Steady state free precession
10) Balanced FFE
11) Multi-echo data image combination
12) Ultrafast gradient echo 2D with preparation pulse
13) Ultrafast gradient echo 3D with preparation pulse
14) Volume-interpolated 3D GRE
15) Aterial Spin Labeling
16) Inversion Recovery
17) Short-Tau IR
18) Long-Tau IR
19) Dual Inversion Recovery
20) Echo Planar Imaging, Diffusion
21) Numer of Echoes in EPI
22) Diffusion-Weighted Imaging
23) Apparent Diffusion Coefficient Map (ADC)
24) Signal uniformity correction for acquisitions made with multi-element coils
25) Diffusion Tensor Imaging
26) High-resolution diffusion imaging
27) Reduced-FoV Imaging for diffusion and BOLD EPI
28) Motion correction algorithm
29) Non-contrast MR Angio
30) Enhances contrast and MTC for reduction of periorbital fat signal.
31) Visualization of blood vessels by phase-sensitive imaging.
32) Background suppression of white and gray brain matter in brain angiography sequences
33) Contrast Bolus Timing/ Visualization
34) Fat saturation
35) Water Excitation
36) Suppression of tissue signal and flow artifacts
37) Enhanced k-space filling reduces venous contamination in 3D high resolution MRA studies
10. Advanced applications
1) Automatic voice audio information to the patinet on length of scan : AutoVoice or Equivalent 1 set
2) Noise reduction during MRI scan : ComforTone or Quite Suite 1 set
3) High resolution and high contrast SWI imaging technique : SWI Specialist or SWI 1 set
4) Visualization of white matter fiber tracts of Diffusion Tensor Imaging (DTI) data 1 set
: FiberTrak Specialist or Syngo DTI Tractography
5) Perfusion imaging : EPI Perfusion or EPI Perfusion Imaging 1 set
6) Quantifying brain perfusion software : 3D ASL or ASL 3D 1 set
7) State-of-the-art acceleration factors in the brain by simultaneously exciting multiple slices 1 set
: MultiBand SENSE or SMS
8) Complete set of single voxel, multi-voxel and multi-slice proton spectroscopy 1 set
: Spectroscopy specialist or Spectroscopy
9) Black Blood imaging without cardiac gating using prepulse (MSDE or equivalent) 1 set
: BlackBlood Imaging or MSDE
10) Special sequences for quantitative assessment of flow 1 set
: Q-Flow Analysis or Flow Quantification
11) 2-point mDIXON technology for TSE sequences : mDIXON XD TSE Specialist or DIXON TSE 1 set
12) Improves soft tissue visualization in the vicinity of MR conditional orthopedic implants 1 set
: O-MAR XD or Advanced WARP
13) Time-Resolved Angiography (using by Keyhole) provides high spatial and temporalresolution 1 set
CE-MRA scans : 4D-TRAK or TWIST
14) Acquisition of multi-slice, dynamic tissue studies : Cardiac Expert Specialist or Advance Cardiac 1 set
15) Diffusion images with small FOV : Zoom DWI Imaging or HyperCUBE & ZOOMit 1 set
16) Produces quantitative fat fraction maps in a single breath-hold, covering the whole liver 1 set
: mDixon Body Fat Quant Specialist or LiverLab
17) Enables automated multi-station head-to-toe coverage 1 set
: WholeBody Specialist or Tim Wholebody Suite
18) Time-resolved 3D technique to drastically accelerate dynamic body and breast imaging 1 set
: 4D-THRIVE and 4D-BLISS or TWIST VIBE
19) Enables a 3D radial FFE acquisition which reduces motion artefacts compared to 3D Cartesian imaging 1 set
during free breathing : 3D Vane XD or Star VIBE
20) Free breathing, motior free Liver dynamic imaging : 4D FreeBreathing or GRASP VIBE 1 set
11. MR Independent Workstation : AVW upgrade to newest version or Syngo Via Server XL 1 set
1) Software Package on workstation (3 Copies for Each Applications) 3 Copies
- Mutil Modality Viewer
- Neuro Perfusion analysis- Mismatch
- T1 Perfusion Analysis
- MobiView
- MR image Subtraction
- cDWI, IVIM, Kurtosis Analysis
- MR Cardiac Function Analysis
- Automa(cid:415)c display of stroke volume, ejec(cid:415)on frac(cid:415)on and end diastolic for both LV & RV
- AI-enabled contours for both LV & RV
- Automatic ED/ES selection
- MR Cardiac Pefusion Analysis
- Polar map display of perfusion (including AHA segments) and parameters
- Automatic contours for Spatial Analysis
- Semi-automatic scar, edema and MVO analysis
- MR Cardiac Delayed Enhancement Analysis or MR Cardiac Spatial Enhancement
12. Site Equipments 1 set
- H.V.A.C 1 set
- Water Chiller 1 set
- Nonmagnetic Patient Safety Camera System (including Out-Room) 1 set
- RF Shield (Existing RF Shield – Partial Modification) 1 set
- RF room interior including skyview coil furniture 1 set
13. Accessory 1 set
- MR Environment Warning Signs 1 set
- Nonmagnetic Auto Injector 1 set
- Nonmagnetic Patient Monitor (MR400) 1 set
- Stretchercart 1 set
- Control Room furniture 1 set
-보조기구(Multipad) 1 set
- Wheelchair for MRI 1 set
14. Unique Items of each vendor : [Type A] or [Type B] 1 set
[Type A]
1) dStream 1 set
2) ScanWise Implant 1 set
3) 2k imaging 1 set
4) dSync (DDAS) 1 set
5) MultiTransmit TX 4D for Cardiac 1 set
6) 2D/3D CSI spectroscopy with SENSE 1 set
7) ABR(Anatomy based regularization) 1 set
8) MAF(Minimal artifact factor) 1 set
9) Multiple time-point ASL for pulse ASL 1 set
10) Patient adaptive B1 calibration for each patient 1 set
11) Advanced motion correction algorithm : MultiVane XD 1 set
12) MutiVane XD with Parallel Imaging technique 1 set
13) 3D NerveVIEW 1 set
14) mDIXON XD for anatomies(Cardiac LDE, Coronary MRA, Subtractionless CE-MRA, …) 1 set
15) mDixon TSE with MultiVane XD for Neuro area 1 set
16) Cardiac Delayed Enhance Analysis 1 set
17) Cardiac Perfusion Analysis 1 set
18) Vital Eye -Camera based Touchless patient sensing 1 set
19) Automatic planning for brain examination : SmartExam Brain 1 set
20) Automatic planning for spine examination : SmartExam Spine 1 set
21) Automatic planning for shoulder examination : SmartExam Shoulder 1 set
22) Automatic planning for knee examination : SmartExam Knee 1 set
23) Automatic planning for breast examination : SmartExam Breast 1 set
24) TSE Diffusion imaging with reduced distortion : Diffusion XD TSE 1 set
25) Time-resolved MRA technology without the use of contrast agent : 4D TRANCE 1 set
26) Improves visualization of the brachial and lumbar plexus by providing you with a high 1 set
resolution T2w TSE acquisition : 3D NerveVIEW 1 set
27) dS Zoom 1 set
28) Enables real-time interactive imaging for planning, while changing geometry and contrast parameters 1 set
during scanning with real-time reconstruction : Real Time Specialist 1 set
29) 3D Turbo Gradient Spin Echo : GRASE 1 set
30) Fat-free FFE imaging with large FOV and sub-millimeter resolution, not only for body imaging, but also for challenging anatomies,
1 set
including head, neck and spine and single station subractionless MRA : mDixon XD FFE Specialist
31) Upgrade adds multi-station imaging for whole body applications : mDixon XD FFE Multi Station 1 set
32) High-resolution istropic 3D TSE acquisition in Brain 1 set
33) High-resolution istropic 3D TSE acquisition in MSK 1 set
34) High-resolution istropic 3D TSE acquisition in Breast 1 set
35) High-resolution istropic 3D TSE acquisition in Spine 1 set
36) Compressed SENSE 2D TSE/FFE/TFE 1 set
37) Compressed SENSE 3D TSE/FFE/TFE 1 set
38) Compressed SENSE 3D MRCP 1 set
39) Compressed SENSE 3D TOF & 3D PCA MR angiogrphy 1 set
40) Compressed SENSE Neuro 1 set
41) Compressed SENSE Spine 1 set
42) Compressed SENSE MSK 1 set
43) Compressed SENSE Breast 1 set
44) Compressed SENSE Upper Body (Liver,kidney, pancreas,Lung) 1 set
45) Compressed SENSE Lower Body (Pelvis, Prostate, Rectum) 1 set
46) Compressed SENSE Cardiac 1 set
47) SmartSpeed 2D TSE/FFE/TFE 1 set
48) SmartSpeed 3D TSE/FFE/TFE 1 set
49) SmartSpeed 3D MRCP 1 set
50) SmartSpeed 3D TOF & 3D PCA MR angiogrphy 1 set
51) SmartSpeed Neuro 1 set
52) SmartSpeed Spine 1 set
53) SmartSpeed MSK 1 set
54) SmartSpeed Breast 1 set
55) SmartSpeed Upper Body (Liver, kidney, pancreas, Lung) 1 set
56) SmartSpeed Lower Body (Pelvis, Prostate, Rectum) 1 set
57) SmartSpeed Cardiac 1 set
58) SmartSpeed Diffusion 1 set
59) SmartSpeed 3D FreeBreathing 1 set
60) SmartSpeed Implant 1 set
61) SmartSpeed MotionFree 1 set
62) SmartSpeed Precise 2D TSE/FFE/TFE 1 set
63) SmartSpeed Precise Neuro 1 set
64) SmartSpeed Precise Spine 1 set
65) SmartSpeed Precise MSK 1 set
66) SmartSpeed Precise Breast 1 set
67) SmartSpeed Precise Upper Body (Liver, kidney, pancreas, Lung) 1 set
68) SmartSpeed Precise Lower Body (Pelvis, Prostate, Rectum) 1 set
69) SmartSpeed Precise Cardiac 1 set
70) Anterior coil frame 1 set
71) Geolink for Multistation exam 1 set
72) Asymmetric profile order 1 set
73) Provide AI upgrade for current system : SmartSpeed & SmartSpeed Precise (Permanent) 1 set
[Type B]
1) Brain Dot Engine 1 set
2) Large Joint Dot Engine 1 set
3) Spine Dot Engine 1 set
4) Abdomen Dot Engine 1 set
5) Cardiac Dot Engine 1 set
6) Multi-shot EPI DWI imaging : RESOLVE 1 set
7) fast RESOLVE 1 set
8) Haste & BLADE Diffusion 1 set
9) High-resolution 3D imaging with reduced breath-hold times : CAIPIRINHA 1 set
10) Ultra shourt TE sequence : PETRA 1 set
11) Reduced-FoV Imaging for 3D TSE : ZOOMit SPACE 1 set
12) Enhances inline BOLD imaging with motion correction during the acquisition of a BOLD exam :3D PACE 1 set
13) Acquire free-breathing, high-resolution Cardiac Cine images : Compressed Sensing Cardiac Cine 1 set
14) RESOLVE with SMS technique, enables scan time reductions as a factor of slice acceleration 1 set
15) Simultaneous Multi Slice TSE & Dixon 1 set
16) Inline composing syngo 1 set
17) Compressed Sensing SEMAC 1 set
18) CAIPIRINHA SPACE 1 set
19) Tim Planning Suite 1 set
20) 4 scan trace diffusion 1 set
21) True FISP and Double Echo steady state : CISS & DESS 1 set
22) A disruptive leap in non-contrast MR Angiography : QISS 1 set
23) TimCT FastVIEW: the “one go” localizer for the whole-body or large body regions such as the whole spine or the whole abdomen 1 set
24) Breast dedicated spectroscopy :Syngo GRACE 1 set
25) NVIDIA GPU for reconstruction 1 set
26) BioMatrix Technology : Respiratory Sensor 1 set
27) BioMatrix Technology : CoilShim 1 set
28) BioMatrix Technology : SliceAdjust 1 set
29) BioMatrix Technology : Select & Go 1 set
30) BioMatrix Technology : Beat Sensor 1 set
31) Motion Insensitive Cardiac Delayed Enhancement : HeartFreeze 1 set
32) Wave CAIPIRINHA SWI 1 set
33) Deep Resolve Gain & Sharp & Boost & Swift Brain 1 set
34) Deep Resolve Diffusion & Haste 1 set
35) Deep Resolve Boost Implant 1 set
36) Quantum 3D Neuro pkg (provide atfer KFDA approval) 1 set
37) Quantum 3D Neurography pkg (provide atfer KFDA approval) 1 set
38) Quantum 3D MSK pkg (provide atfer KFDA approval) 1 set
39) Quantum 3D Abdomen pkg (provide atfer KFDA approval) 1 set
40) Quantum 3D Breast pkg (provide atfer KFDA approval) 1 set
41) Turbo Flash Mode (Ultra fast protocol) 1 set
42) Provide AI upgrade for current system : SwiftMR(5 years) 1 set
E. REMARKS
1.Installation : Contractor is responsible for installation of this equipment at customer's premise.
1) 장비가 병원도착 후 원내 반입 및 설치완료시 까지 일체비용은 계약업체 또는 공급사가 부담하여야 한다.
2) 입찰(규격)등록 시 : 발주기관과 협의하여 시방서와 평면도면을 제출하여야 한다.
3) 계약체결 시: 계약 체결 시점 혹은 계약후 15 일이내에 시설공사에 대한 상세도면, 시방서, 설치계획서(일정) 등을
발주기관과 협의하여 제출하여야 한다.
2.Warranty: Contractor should be responsible for free of charge after service whenever there is any defect or trouble for 3
years after operation test. (Including all parts)
1) 3 년 무상 보증 기간 동안 연 가동률 95 %이상으로 보장해야 한다. 그렇지 않은 경우, 공급 업체는 가동 중지 기간과 같은 무
상 보증 기간 연장을 제공해야 한다.
3.Contractor or supplier shall provide the one set of unique items among [A type] or [B type]
4.Maintenance service
1) Contractor or supplier shall provide factory-trained service persons who are staying in Korea for after-installation
service in any situation and should preserve spare parts.
2) The contractor will provide continuous updates to the latest S/W version under warranty/maintenance contract.
3) The contractor is responsible for replacement of H/W system, if required to operate the latest S/W under
warranty/maintenance contract.
5.Contractor or Supplier should provide Service Key to customer.
6.병원 요청 시 국내교육을 실시하여야 한다.
7.제조 1년 이내의 제품을 납품하여야 한다.
8.하자보수기간 종료 전 병원 측의 장비 최종점검 요청에 협조하여야 한다.
9.공급자는 하자보증기간내 장비 안정성을 위하여 병원측의 예방점검 요청에 적극 협조하여야 한다.
10.납품 이후 1 년 이내 단종 예정이 없는 제품을 납품해야 한다.
11.장비 설치 시 공급자는 PACS 서비스에 대해 무상 지원해야 한다.
12.장비 정기 점검은 격달 1회를 기본으로 하되 사용자 요구시 추가나 변경 진행할 수 있다.