전자동 NGS 전처리 시스템 구입
| 공고기관 | 경북대학교병원 |
| 수요기관 | 경북대학교병원 |
| 입찰방식 | 전자입찰 |
| 계약방법 | 일반경쟁 |
| 예가방식 | 단일예가 |
| 낙찰방식 | 최저가낙찰제-최저가낙찰자 |
| 배정예산 | 630,000,000 |
| 공고일시 | 2026-10-06 13:31 |
| 마감일시 | 2026-10-14 10:00 |
| 개찰일시 | 2026-10-14 11:00 |
| 참가신청마감 | - |
| 담당자 | 김병문 |
| 담당자 전화 | 053-200-5048 |
| 담당자 이메일 | iopblack789@naver.com |
참가자격
| 참가가능지역 | 전국 |
| 업종 | 의료/약품 |
| 경쟁방식 | 일반경쟁 |
| 공동수급 | (없음)공동수급불허 |
COMMODITY DESCRIPTION
| H.S. NO | ITEM NO. | DESCRIPTION | QUANTITY | UNIT |
| Automated NGS preparation system | 1 | System |
A . FEATURES :
1. The system provides fully automated sample preparation for next-generation sequencing (NGS)
workflows.
2. The system supports automated liquid handling for nucleic acid extraction, library preparation,
purification, and normalization.
3. The system enables high-throughput processing with reproducible and standardized sample
preparation.
4. The system supports flexible workflow configuration compatible with a wide range of NGS library
preparation protocols.
5. The system provides barcode-based sample tracking and workflow management to ensure traceability
and data integrity.
6. The system minimizes manual intervention to improve laboratory efficiency, reproducibility, and
operational reliability.
B. CONSIST OF :
1. Main body 1 set
2. Accessories 1 set
C. SPECIFICATIONS :
1. Main body
1) Pipetting channels : 96-channel or more
2) Pipetting type : syringe-based pipetting system or
equivalent
3) Partial tip loading : 1 ~ 96 tips or more
4) Normalization : Supports automatic normalization
based on user-defined
concentration values or equivalent
5) Deck positions : 12 or less
6) SPRI cleanup and double SPRI size selection function
7) PCR setup and normalization function
Page 1 of 3 신청부서 : 병리과 의공기술검토자: 나승대 (인)
8) Plate lid handling : Integrated gripper for automated
plate lid removal and replacement
or equivalent
9) 5-Axis robotic arm function
10) External tip storage : 27 tip racks or more
11) Safety : Equipped with a manufacturer-
provided enclosure and emergency
stop button or equivalent
12) Labware movement : Integrated gripper for automated
movement of various labware or
equivalent
13) Thermal block
a) Number of Blocks Provided : 1 or more
b) Temperature range : 4 ~ 110 degree or more
14) Thermal shaker
a) Number of Blocks Provided : 1 or more
b) Temperature range : 4 ~ 70 degree or more
15) Number of fully automated PCR systems provided : 1 or more
16) Tip management : Provides both internal tip waste
and external tip box waste
positions or equivalent
17) Emergency stop : Equipped with an emergency stop
button for immediate interruption of
system operation or equivalent
18) Deck check system : Vision-based deck check system
for automatic verification of
labware type, deck position, and
reagent volume prior to operation
or equivalent
19) Throughput : 96 samples or more
20) UV lamp : Equipped with a UV lamp for
contamination control or equivalent
2. Accessory
1) Operating system 1 ea
2) Compressor 1 ea
3) Support application development (S/W) 1 ea
Page 2 of 3 신청부서 : 병리과 의공기술검토자: 나승대 (인)
4) Table 1 ea
5) UPS system 1 ea
6) Operation manual 1 ea
7) Service manual 1 ea
D. REMARKS :
1. Installation
- Supplier or contractor should be responsible for installation of this equipment.
2. Warranty
- Supplier or contractor should be responsible for free of charge after - service whatever there is any
defect or any trouble for three years after being accepted by customer.
- If contractor and supplier do not supply appropriate service manual, the period of time for warranty is six
years.
- If a warranty of 3 years or less is provided for all or part of the equipment, it is separately described in
section B. Consist of.
3. Etc.
- If there is a request from the user department, the supplier provides hospital computer data linkage (1-
year warranty) such as EMR, PACS, and LIS free of charge, and the procedures of data link should be
proceeded with the department of medical information in KYUNGPOOK NATIONAL UNIVERSITY
HOSPITAL.
- The operating system required for operating the equipment should be based on Win11Pro. In case of
lower versions, the supplier must report to the KNUH Information Security Team and should perform
free upgrades of the operating system and operating software according to future equipment updates.
Page 3 of 3 신청부서 : 병리과 의공기술검토자: 나승대 (인)